25-apr-11 22:10:00 Measurements ended --- . Number of lamps (L) = 7 . Filter running time = 52.24 hours . Total airflow this filter = 15.33 cubic meters . Mean BC concentration of all lamps = 89 ng/m3 . Report for lamp 1 (370 nm): . Ending voltages: sen(beam) = .497 sen(zero) = 0.021 . ref(beam) = 2.250 ref(zero) = 0.021 . Optical attenuation of filter deposit = 99 . Total aerosol black carbon on filter = 1256 ng Mean aerosol black carbon concentration = 82 ng/m3 --- The reference beam showed lamp intensity fluctuations of 54 ppm. --- Report for lamp 2 (470 nm): . Ending voltages: sen(beam) = .925 sen(zero) = 0.021 . ref(beam) = 2.448 ref(zero) = 0.021 . Optical attenuation of filter deposit = 83 . Total aerosol black carbon on filter = 1342 ng Mean aerosol black carbon concentration = 87 ng/m3 --- The reference beam showed lamp intensity fluctuations of 42 ppm. --- Report for lamp 3 (520 nm): . Ending voltages: sen(beam) = .676 sen(zero) = 0.021 . ref(beam) = 2.523 ref(zero) = 0.021 . Optical attenuation of filter deposit = 74 . Total aerosol black carbon on filter = 1318 ng Mean aerosol black carbon concentration = 86 ng/m3 --- The reference beam showed lamp intensity fluctuations of 34 ppm. --- Report for lamp 4 (590 nm): . Ending voltages: sen(beam) = .951 sen(zero) = 0.021 . ref(beam) = 1.226 ref(zero) = 0.021 . Optical attenuation of filter deposit = 67 . Total aerosol black carbon on filter = 1359 ng Mean aerosol black carbon concentration = 88 ng/m3 --- The reference beam showed lamp intensity fluctuations of 142 ppm. --- Report for lamp 5 (660 nm): . Ending voltages: sen(beam) = 1.377 sen(zero) = 0.021 . ref(beam) = 2.522 ref(zero) = 0.021 . Optical attenuation of filter deposit = 62 . Total aerosol black carbon on filter = 1404 ng Mean aerosol black carbon concentration = 91 ng/m3 --- The reference beam showed lamp intensity fluctuations of 89 ppm. --- Report for lamp 6 (880 nm): . Ending voltages: sen(beam) = 1.427 sen(zero) = 0.021 . ref(beam) = 3.544 ref(zero) = 0.021 . Optical attenuation of filter deposit = 47 . Total aerosol black carbon on filter = 1440 ng Mean aerosol black carbon concentration = 94 ng/m3 --- The reference beam showed lamp intensity fluctuations of 95 ppm. --- Report for lamp 7 (950 nm): . Ending voltages: sen(beam) = 1.168 sen(zero) = 0.021 . ref(beam) = 4.555 ref(zero) = 0.021 . Optical attenuation of filter deposit = 43 . Total aerosol black carbon on filter = 1417 ng Mean aerosol black carbon concentration = 92 ng/m3 --- The reference beam showed lamp intensity fluctuations of 97 ppm. --- Disk space = 1.873408E+07 bytes free: One entry at every 5 minutes. There is space for 3983 hours ( 166 days) more data --- 25-apr-11 22:10:00 Tape feeder mechanism advancing for 1 spot(s). --- ============================================================= --- MAGEE SCIEN. AETHALOMETER --- Aethalometer No. 463 --- 25-apr-11 22:24:54 Measurements started Starting voltages: . sen(zero) = 0.021 ref(zero) = 0.021 . Lamp 1 (370 nm): sen(beam) = 1.309 ref(beam) = 2.237 . Lamp 2 (470 nm): sen(beam) = 2.108 ref(beam) = 2.435 . Lamp 3 (520 nm): sen(beam) = 1.403 ref(beam) = 2.530 . Lamp 4 (590 nm): sen(beam) = 1.844 ref(beam) = 1.219 . Lamp 5 (660 nm): sen(beam) = 2.541 ref(beam) = 2.498 . Lamp 6 (880 nm): sen(beam) = 2.306 ref(beam) = 3.536 . Lamp 7 (950 nm): sen(beam) = 1.789 ref(beam) = 4.526 ---