NEUBrew Network Instrument Calibrations/Characterizations


Ozone Calibrations

Brewer UID
Calibration Date
Notes
BR101 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR105 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR108 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR131 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR131 2013-05-26 Ozone Calibrarion performed by Volodya Savastiouk of International Ozone Services (IOS)
BR131 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR132 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR134 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR134 2013-05-26 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR134 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR135 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR139 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR139 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR140 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR140 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR141 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR141 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR144 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR144 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR146 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR146 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR147 2006-08-20 Ozone Calibration performed by Ken Lamb from IOS.
BR147 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
BR154 2007-02-28 Ozone Calibration performed by Ken Lamb from IOS.
BR154 2013-05-26 Ozone Calibrarion performed by Volodya Savastiouk of International Ozone Services (IOS)
BR154 2014-10-19 Ozone Calibration performed by Volodya Savastiouk of International Ozone Services (IOS)
Record Count: 26

UV-B Calibrations


Calib UID
Brewer UID
Calibration Date
Notes
90 BR108 2015-10-22 UB-B Calibration by Patrick Disterhoft. Lamp used: H-070
30 BR134 2007-04-18 Spectral Responsivity calibration performed by P Disterhoft. Lamps used were H-025 and H-027 in diffuse and total signal.
41 BR134 2008-08-21 UV-B Calibration performed by Patrick Disterhoft from CUCF. Lamps used H-041 and H-027
49 BR134 2010-09-22 UV-B Calibration performed by Patrick Disterhoft.
84 BR134 2015-04-07 Calibration Performed by Patrick Disterhoft. Lamps:H-062 & H-070
92 BR134 2015-11-19 Calibration performed by Patrick Disterhoft. Lamps run: H-062 & H-070
96 BR134 2016-06-30 Calibration performed by Patrick Disterhoft. Lamps run: H-070
99 BR134 2017-06-29 Calibration performed by Patrick Disterhoft. Lamps run: H-062
111 BR134 2020-10-29 Calibration performed by Scott Stierle. Lamps: H-066, H-077, H-078
114 BR134 2021-07-15 UV Calibration by Scot Stierle using H-066,H-077-H-078. H-066 was not used in the Final Calibration applied.
117 BR134 2022-08-23 UV Calibration by Scott Stierle using Lamps H-066, H-077, H-078.
113 BR135 2021-07-08 Calib by SWS using lamps: H-066, H-077, H-078
31 BR139 2007-04-18 Spectral Responsivity calibration performed by P Disterhoft. Lamps used were H-025 and H-027 in diffuse and total signal measurements. UV Cal not valid after 2007-08-09 when nickel sulfide filter replaced by UG-11x filter.
39 BR139 2007-08-10 Artificial UV responsivity generated from CI Scan Ratios after the Nickle Suphate Filter swapped out for UG-11.
40 BR139 2008-02-12 Artificial UV responsivity generated from CI Scan Ratios after an single UG-11 Filter swapped out for back to back UG-11 filters.
85 BR139 2015-04-07 Calibration Performed by Patrick Disterhoft. Lamps:H-062 & H-070
95 BR139 2015-11-19 Calibration performed by Patrick Disterhoft. Lamps run: H-070
97 BR139 2016-06-30 Calibration performed by Patrick Disterhoft. Lamps run: H-070
100 BR139 2017-06-29 Calibration performed by Patrick Disterhoft. Lamps run: H-062
110 BR139 2020-10-21 Calibration performed by Scott Stierle. Lamps: H-066, H-077, H-078
115 BR139 2021-07-16 UV Calibration by Scott Stierle using Lamps H-066, H-077, H-078. All three lamps used in calibration calculations.
118 BR139 2022-08-29 UV Calibration by Scott Stierle using H-066, H-077, H-078. H-078 irradiance was not used in responsivity averaging.
17 BR140 2006-10-13 UV-B Calibration performed by Patrick Disterhoft after Brewer installed at Raleigh, NC.
36 BR140 2007-10-04 UV-B Calibration performed by Patrick Disterhoft at Raleigh, NC.
51 BR140 2010-10-13 UV-B Calibration performed by Patrick Disterhoft.
54 BR140 2011-10-06 UV-B Calibration performed by Patrick Disterhoft.
80 BR140 2014-11-17 UV Calibration performed by Patrick Disterhoft. Lamps: H062/H066
89 BR140 2015-10-15 UB-B Calibration by Patrick Disterhoft. Lamps used: H-062, H-070
105 BR140 2018-10-18 Calibration performed by Patrick Disterhoft. Lamps: H066,H077,H078
121 BR140 2021-11-09 UV Calibration by Scott Stierle using H-066, H-077, H-078. H-077 was not used in responsivity Avg calculations.
32 BR141 2007-04-18 Spectral Responsivity calibration performed by P Disterhoft. Lamp sequence:H-041, H-025, H-027; both diffuse and total scan measurements.
42 BR141 2008-08-21 UV-B Calibration performed by Patrick Disterhoft from CUCF. Lamps used H-041 and H-027
46 BR141 2009-03-19 UV-B Calibration performed by Patrick Disterhoft from CUCF. Lamp used H-041
86 BR141 2015-04-07 Calibration Performed by Patrick Disterhoft. Lamps:H-062 & H-070
93 BR141 2015-11-19 Calibration performed by Patrick Disterhoft. Lamps run: H-070
98 BR141 2016-06-30 Calibration performed by Patrick Disterhoft. Lamps run: H-070
101 BR141 2017-06-29 Calibration performed by Patrick Disterhoft. Lamps run: H-062
109 BR141 2020-10-15 Calibration performed by Scott Stierle. Lamps: H-066, H-077, H-078
116 BR141 2021-07-16 UV Calibration by Scott Stierle using lamps H-066, H-077, H-078. H-078 was not included in the Responsivity Averaging Calculations.
119 BR141 2022-08-24 UV Calibration by Scott Stierle using H-066, H-077, H-078. H-077 not used on responsivity Average calculations.
15 BR144 2006-09-24 UV-B Calibration performed by Patrick Disterhoft after Brewer was installed at the Bondville, IL.
33 BR144 2007-06-20 UV-B Calibration performed by Patrick Disterhoft.
43 BR144 2008-08-29 UV Calibration Performed by Patrick Disterhoft. Lamps used H038, H027, H041.
52 BR144 2011-09-14 UV-B Calibration performed by Patrick Disterhoft.
81 BR144 2014-11-21 UV Calibration Performed by Patrick Disterhoft. Lamps: H062/H070
87 BR144 2015-08-25 UV Calibration performed by Patrick Disterhoft. Lamps: H-062,H-066,H-070.
108 BR144 2018-09-13 Calibration Performed by Patrick Disterhoft. Lamps: H066,H077,H078
122 BR144 2021-11-04 UV Calibration by Scott Stierle using lamps H-066, H-077, H-078.
24 BR146 2007-03-27 UV-B Calibration performed by Patrick Disterhoft from CUCF.
37 BR146 2007-11-06 Calibration performed by Patrick Disterhoft at the MRS.
38 BR146 2008-02-21 UV-B Calibration Performed by Patrick Disterhoft
47 BR146 2009-04-08 UV-B Calibration performed by Patrick Disterhoft from CUCF. Lamp used H-041
48 BR146 2010-09-09 UV-B Calibration performed by Patrick Disterhoft.
53 BR146 2011-09-26 UV-B Calibration performed by Patrick Disterhoft.
82 BR146 2014-10-30 UV Calibration performed by Patrick Disterhoft. Lamps: H062/H066/H070
94 BR146 2015-11-23 Calibration performed by Patrick Disterhoft. Lamps run: H-066 & H-070
107 BR146 2018-11-07 Calibration performed by Patrick Disterhoft. Lamps: H066,H077,H078
123 BR146 2022-10-05 UV Calibration by Scott Stierle using lamps H-066, H-077, H-078.
20 BR147 2006-11-06 UV-B Calibration performed by Patrick Disterhoft after Brewer was installed at Ft. Peck, MT.
34 BR147 2007-09-27 UV-B Calibration performed by Patrick Disterhoft Brewer Temperature 25.4 C.
50 BR147 2010-09-29 UV-B Calibration performed by Scott Stierle.
55 BR147 2011-10-21 UV-B Calibration performed by Patrick Disterhoft.
58 BR147 2012-08-22 UV-B Calibration performed by Patrick Disterhoft. For the August 22, 2012 calibration, the lamps were run in the following order: H-060, H-041, H-058, H-059. Their irradiances were determined during the CUCF scale transfer #76. The average temperature of the Brewer during the four lamp calibrations was 32.81 degrees C.
59 BR147 2013-10-03 UV-B Calibration performed by Patrick Disterhoft. For the October 3, 2013 calibration, the lamps were run in the following order: H-058, H-059 and H-041. Their irradiances were determined during the CUCF scale transfer #87. The average temperature of the Brewer during the three lamp calibrations was 14.01 degrees C.
83 BR147 2014-11-05 UV calibration performed by Patrick Disterhoft. Lamps: H062/H070
88 BR147 2015-09-23 UV calibration performed by Scott Stierle. Lamps: H-062,H-066,H-070.
106 BR147 2018-09-26 Calibration performed by Patrick Disterhoft. Lamps: H066,H077,H078
112 BR147 2019-09-14 Calibration performed by Scott Stierle. Lamps: H-066, H-077, H-078
124 BR147 2021-09-08 UV Calibration by Scott Stierle using lamps H-066, H-077, H-078.
125 BR147 2022-10-05 UV Calibration by Scott Stierle using lamps H-066, H-077, H-078.
13 BR154 2006-07-21 UV-B Calibration performed by Patrick Disterhoft after Brewer was installed at Houston, TX.
23 BR154 2007-03-01 UV-B Calibration performed by Patrick Disterhoft from CUCF.
56 BR154 2011-12-09 UV-B Calibration performed by Patrick Disterhoft.
57 BR154 2013-06-06 UV-B Calibration performed by Patrick Disterhoft.
79 BR154 2014-11-12 UV Calibration performed by Patrick Disterhoft. Lamps used: H062/H066/H07
102 BR154 2016-05-02 Calibration perfromed by Patrick Disterhoft. Lamps Run: H-062, H-066, H-070
103 BR154 2017-02-23 Calibration perfromed by Patrick Disterhoft. Lamps Run: H-066, H-070
104 BR154 2018-08-02 Calibration performed by Patrick Disterhoft. Lamps Run: H-066 and H-067.
Record Count: 78

Cosine Measurements

Brewer UID
Calibration Date
Notes
BR134 2007-02-15 Cosine Measurement performed by Patrick Disterhoft at DSRC.
BR139 2007-01-24 Cosine Measurement performed by Patrick Disterhoft at DSRC.
BR140 2006-10-01 Cosine Measurement performed by Patrick Disterhoft at DSRC prior to the Brewer being packed up and shipped to the Raleigh NC site.
BR141 2007-03-12 Cosine Measurement performed by Patrick Disterhoft at DSRC.
BR144 2006-09-18 Cosine Measurement performed by Patrick Disterhoft at DSRC.
BR146 2006-10-09 Cosine Measurement performed by Patrick Disterhoft at DSRC.
BR147 2006-10-25 Cosine Measurement performed by Patrick Disterhoft at DSRC.
BR154 2006-07-15 Cosine Measurement performed by Patrick Disterhoft at DSRC prior to deployment to Houston, TX.
Record Count: 8

Spectral Lamp Measurements

Brewer UID
Calibration Date
Notes
BR134 2007-02-01 External HG,CD,ZN Lamps Scanned; Brewer internal HG Lamp Scanned. Measurments made at DSRC.
BR139 2007-01-25 External HG,CD,ZN Lamps Scanned; Brewer internal HG Lamp Scanned. Measurments made at DSRC.
Record Count: 2