A Low-Maintenance Drying System for Ambient Air Greenhouse Gas Monitoring
L.R. Welp1, W. Paplawsky1, R.F. Keeling1, R.F. Weiss1 and S. Heckman2
1Scripps Institution of Oceanography (SIO), University of California at San Diego, La Jolla, CA 92093; 858-822-1642, E-mail: lwelp@ucsd.edu
2Earth Networks, Inc., Germantown, MD 20876
In preparation for the routine deployment of the Earth Networks greenhouse gas monitoring network, we have designed and tested a simple method for drying ambient air to below 0.1% mole fraction H2O using a Nafion model MD-050-72S-1 dryer. We tested the prototype inlet drying system on a Picarro model G2301 Cavity Ring-Down Spectroscoy CO2/CH4/H2O analyzer. The analyzer measures water vapor mixing ratio at the same frequency as CO2 and CH4 and then corrects for the dilution and peak broadening effects on the CO2 and CH4 mixing ratios. This analyzer is remarkably stable and performs well on water vapor correction tests, but we believe there is an added benefit of reducing the dependence on the H2O correction for long-term field measurement programs. Unlike CO2 and CH4 measurements, which can be calibrated in the field with compressed gas standards, the H2O correction is not easily calibrated in the field. Substantially lowering the amount of H2O in the sample can reduce uncertainties in the applied H2O corrections by an order of magnitude or more, and reduce the need to verify the H2O correction algorithm stability.
Our Nafion drying inlet system takes advantage of the extra capacity of the external Picarro analyzer pump to redirect 30% of the dry gas exiting the Nafion to the outer shell side of the dryer. This method has no consumables and does not appear to alter the CO2 and CH4 concentrations of the sample gas within measurement precision. We restrict sample flow through the G2301 analyzer to 70 cc/min to conserve calibration gases during routine deployment. It is difficult to humidify air without changing its dry-gas CO2 mixing ratio slightly because of the propensity of CO2 to interact with water adsorbed on tubing walls. For this reason, we compared our Nafion dryer system with a proven cold trap at -95°C. Dry air from a standard tank was passed through a humidifier in the wet-gas test (or bypassed in the dry-gas test) and then alternately directed through the Nafion or cold trap at 15-min intervals. Differences between the Nafion dryer and cold trap in the dry-gas and wet-gas tests were less than the targeted precision of the measurements. Systematic differences between the drying methods were at the level of 0.01 ppm in CO2 and 0.007 ppb in CH4 for the wet-gas test (Fig 1) and 0.01 ppm in CO2 and 0.04 ppb in CH4 for the drygas test.